Geophysics Department TU Clausthal
|
Applied Geophysics : Gravity
|
A Gravity Meter
Applet
( in separate window, ca. 440 x 530 Pixel )
- The applet simulates a Lacoste-Romberg Mod D Gravity Meter,
- described in detail in
L&R MODEL G AND D METER MANUAL
- ( PDF, courtesy of
L and R Meter Service, LLC ).
- It shows
- a simplified representation of the "black lid" to operate
the instrument via mouse clicks and to observe its reactions
- and
- a menue panel ( background orange ) to contol the mode of
operation and the program flow.
- The simulation covers the following modes of operation :
- test and adjustment of the instrument
( LABORATORY TEST ),
- calibration of the instrument
( CALIBRATE OUTPUT / FEEDBACK / COUNTER ),
- 2 field surveys
( NETTLETON / MODEL PROFILE ),
- and
- a demonstation of cross coupling
( CROSS COUPLING ).
Table of Content
- Comments
- HowTo
- "Black Lid"
( Gravity Meter )
- Menue Panel ( Program Control )
- Window HELP ( Help / Information )
- Window READINGS ( Display /
Processing )
- Saving and Processing of Readings
- Window CHART RECORDER
( "Analog" Recorder )
- Window CROSS COUPLING ( Cross Coupling
Demo )
- Examples ( Display / Interpretation of
readings )
- Download
Comments
The position of the beam is calculated recursively in constant time
steps of 50 [msec] ( = 20 steps/second ), based on
- a "zero length" spring in Lacoste geometry and with
adjustable upper suspension point,
- see diagram in PRIMARY INFORMATION of
L&R MODEL G AND D METER MANUAL,
- Lacoste geometry :
- see also
LP Seismometer Applet,
( HELP + INFO ! enabled ! ),
- an "ideal" lever system, adjustment screw and gear box
- => uniform calibration factor, readings independent of direction
of nulling,
- an external force, corresponding to the station gravity
( + tidal variations + seismic noise ),
- and split up into
- a "vertical" component ( to be observed ) and
- a "horizontal" component ( parallel to
the beam, => sensitivity change ),
- some details ( mass, momment of inertia, spring constant ) not
known and therefore replaced by estimated values :
- a natural period of 12 [sec] ( damping approx.
0.7 critical ) and
- a distance of approx. 350 [seca] ( seconds of arc ) from
lower to upper stop ( corresponding to approx.
1.4 [mgal] ),
- a feedback force derived from the actual excursion of the beam and
its velocity,
- with ( possibly not very realistic ) coefficients that reduce
the natural period to approx. 0.6 [sec], preserving the damping.
- The real time behaviour of the applet depends on the local computing
power, and
- the link above starts the applet updating the graphic
display every approx. 50 [msec]
( = time step of calculation ).
- Preserving the time step of simulation ( 50 [msec] ),
- a
Fast Version
of the same applet ( approx. 25 [msec] per graphic
frame ) can be startet, and
- for slower computers a
Slow Version
( approx. 100 [msec] per frame ) is implemented.
- The effective frame rate is checked every 5 [sec], and
- the actal value is listed to the screen or the JAVA Console,
if HELP + INFO ( window HELP ) are
activated.
Table of Content
HowTo
"Black Lid" :
( Gravity Meter, screenshot )
- BEAM POSITION :
- Microscope eyepiece to observe the actual beam position ( red
crosshair ) on a scale with reading line
( blue mark ).
- The central part of the scale surrounding the reading line
and covering the interval from lower to upper mechanical stop of the beam
( approx. 14 minior scale divisions =
approx. 1.4 [mgal] )
is "zoomed" to improve the resolution of reading.
- The reading line is preset and corresponds to a preset reference
line of the long level
( L LEV = 0 [seca], green ).
- In LABORATORY TEST ( see below ) a mouse click onto the
red crosshair generates a new reading line at the
corresponding position of the scale.
- ( i.e. after altering the mechanical sensitivity,
see CHECKING THE READING LINE in
L&R MODEL G AND D METER MANUAL ).
- FEEDBACK :
- On/Off toggle switch for the feedback system
- ( ON = red / OFF = blue ).
- M E T E R :
- Frees / Clamps the mechanical system and inhibits / enables
program or station changes ( transportation of the
instrument ! ).
- ( F R E E = red / LOCKED = blue ).
- OUTPUT VOLTAGE :
- Two digital voltage meters display the beam position voltage and
the feedback voltage.
- BEAM POS. [mV]
- beam position relative to reference line
( red T mark = 0 [mV] ) on enlarged scale
( scaling factor approx. 1 [mgal/V] ).
- FEEDBACK [mV]
- feedback voltage ( = BEAM POS.
x 400, scaling factor approx. 1 [mgal/V] for closed loop =
FEEDBACK ON ).
- ADJ ( enabled in LABORATORY TEST )
- selects a new reference, common to both voltages,
( = nulling the electric circuits ) :
- reading line ( normal ) or actual beam postion.
- Both output voltages are low pass filtered :
- DIR = 2-pol Bessel, 0.5 [sec],
- AVR = 6-pol Bessel, 48 [sec].
- After a manual or automatic reset of the AVR low pass
to new initial values
- the output display changes to cyan for approx. 90 [sec],
to avoid incorrect readings.
- SPRING ADJUST ( DIAL ) :
-
- RANGE = red
- +/- => approx. 7 [mgal],
++/-- => 70 [mgal],
+++/--- => 700 [mgal],
- FINE = green
- combines Nulling Dial and Counter,
- approx. 1 [mgal/Unit] => resolution
approx. 1 [µgal],
- FREE ( red ) / LOCK ( blue )
- frees / locks DIAL + RANGE
- L / X LEV :
- The bubble levels have different sensitivities and the "pitch"
of the leveling screws is chosen to produce one screen pixel of bubble
movement for one increment of the corresponding leveling screw.
- Long screw ( right )
- +/- = approx. 1 [µm] =>
approx. 1 [seca] on L LEV,
- Cross screws ( left )
- +/- = approx. 2 [µm] =>
approx. 2 [seca] on X LEV,
approx. 1 [seca] on L LEV,
- L + X screws
- ++/-- = x 10,
+++/--- = x 100.
- L LEV mark ( green )
- is preset to a postion corresponding to the actual reading line, and
- is resulting in a mechanical sensitivity of
approx. 250 [seca/mgal]
- ( => 1 [mgal =
approx. 10 minior divisions on the optical beam
position scale ),
- X LEV mark ( green )
- should be repositioned by moving the bubble to the desired postion and
- clicking on the red crosshair after performing the Cross Level Check
- ( see CHECKING THE TRANSVERSE LEVEL in
L&R MODEL G AND D METER MANUAL ).
- Changing the meter parameters,
- i.e setting a new reading line,
a new output reference or new L LEV marks,
- ! in
LABORATORY TEST or CROSS COUPLING only !
"Black Lid"
Table of Content
Menue Panel :
( Program Control, screenshots )
Inititial state after start of applet
( all menue items disabled, except I N I T )
After I N I T ( button label changed to RESET )
- RUN :
- Starts continuous calculation and display
( 20 [steps/sec] ).
- HALT :
- "Freezes" the actual state.
- STEP :
- Single time step + HALT.
- RESET :
- Resets the gravity meter to initial settings and erases saved readings of
the actually selected procedure.
- ! If activated in LABORATORY TEST all parameters
( calibration factors etc. ) previously determined are reset to
their initial ( default ) values !
- HELP :
- Generates a separate window to display and set various parameters, and
- to list informations ( subsurface model, station table, saved /
processed readings etc. ,
see Window HELP ).
- NO / LO / HI SEIS :
- Sets the amplitude of seismic noise for LAB TEST,
CALIB ... and ... PROFILE
- ( vertical component only )
- NO = no noise,
LO = 3 [µgal],
HI = 30 [µgal],
- ( period 4.8 [sec], beat 72 [sec],
=> vertical ground displacement of 0, 17, 170 [nm] ).
- RES AVR :
- Resets the AVR low pass to new initial values
- ( and changes color of corresponding display / recorder trace to cyan
for approx. 90 [sec] ).
- CHART RECORDER :
- Connects a 2-channel "analog" recorder to the
instrument
- ( see Window CHART RECORDER ).
- MODE OF OPER :
- Program selection to perform tests, calibrations and field surveys
or to run the cross coupling demo.
- LABORATORY TEST
- test and adjustment procedures
- (see CHECKING LEVELS AND SENSITIVITY in
L&R MODEL G AND D METER MANUAL ).
- "Adjustment" of the cross level
( X LEV ADJUST ) by fitting a polynomial
of second degree to readings taken with different cross level settings.
- CALIBRATE OUTPUT / FEEDBACK / COUNTER
- determination of calibration factors ( [mgal/V] ) for
- the output voltages
BEAM POS. [mV] and FEEDBACK [mV],
- and for
- the DIAL / COUNTER combination ( [mgal/Count )
- by fitting a straight line to readings taken along calibration profiles
- ( stations with known gravity values ).
- NETTLETON PROFILE
- determination of the density of a homogenious halfspace from field
observations along a profile with significant topogaphy.
- ( NETTLETON : calculation of cross correlation between obseved
Bouguer anomaly and topography for different, user adjustable density
values,
- PARASNIS : fitting a straight line to observed free air
anomaly versus station height. )
- MODEL PROFILE
- determination of the density of the top layer of a 2-layer model
from field observations along a profile over a brigde across a railway
cutting.
- Fitting a model curve to observed gravity values by variing the layer
density.
- ! Model calulations are 2 D
( Talwani ) !
- CROSS COUPLING
- demonstration of the cross coupling effect produced by coherent
"harmonic" vertical and horizontal accelerations
( shipborne gravity surveys ).
- SAVE STN READING :
- Saves a meter reading for graphical display, processing and interpretation
in a separate frame
( see Window READINGS ).
- NEXT / PREV / BAS
- "Moves" the ( ! locked ! ) instrument
to the next, previous or base station.
Menue Panel
HowTo
Table of Content
Window HELP :
( Help / Informations, screenshot )
- Menue panel to display and modify parameters and the program
environment.
- For the parameters to be determined in LABORATORY TEST
and CALIBRATE OUTPUT, FEEDBK and COUNTER,
- one of three "revisions" of each parameter value
can be displayed :
- the value currently used ( ACTUAL VAL. ),
- the value determined / set by the user
( USER CALIB. )
- and
- the value used in forward calulation of the respective reading
value / station gravity ( "TRUE" VALUE ).
- SET sets ACTUAL VAL. ( the parameter value to be
used )
- to USER CALIB. or
- to "TRUE" VALUE
( enabling the user to skip the corresponding calibration
procdure ).
- The "TRUE" VALUE for the output voltage
BEAM POS. [mV] refers to the preset values of
L LEV ZERO and reading line,
- i.e. BEAM POS. [mV] needs to be recalibrated after
a change in mechanical sensitivity.
- ( X LEV ZERO and counter calibration
- should be independent of mechanical sensitivity,
- feedback calibration
- should be affected by
approx. 1/400. = 1/feedback_gain,
- negligible for nulling FEEDBACK [mv] to values below
approx. ±10 [mV]. )
- METER TILT : RANDOM / CONST. toggels between
- a randomly set tilt of the instrument, needing a new
horizontation after each change of observation station
( ! realistic ! ) and
- a constant tilt after station / program change
( unrealistic, but greatly simplifying the use ).
- WARN inhibits
- saving of incorrect readings ( beam still moving, incorrect
levelling, etc. ) and
- starting field observations with incorrect calibration factors
( [mgal/V] or [mgal/Count] ).
- INFO and HELP ( active in HALT mode only )
display
- informations to the the actual cursor position ( INFO ) and
- hints to actually possible mouse interactions ( HELP )
- in the "Black Lid" and in the window READINGS.
- AUTO CALIB is implemented for impatient users,
- skips test and
calibration procedures and sets
- the calibration parameters to their "true"
values and
- the METER TILT to CONSTANT.
- In the text area informations, concerning the program actually
running, are displayed :
- LIST ... lists the model parameters, the station table and saved
readings,
- TRACE READINGS traces the saving of readings,
- CLEAR erases the text area.
- HIDE hides the window.
Window HELP
HowTo
Table of Content
Window READINGS :
( Display / Processing, screenshot )
Example = raw readings of NETTLETON PROFILE
( more examples )
Display, processing and interpretation of meter readings, display of the
subsurface model.
- A gravity reading, saved via SAVE STN READING
includes :
- the station number ( the cross level reading [seca]
for X LEV ADJUST ),
- the observation time ( [min] ),
- the dial / counter reading ( [Counts] ),
- the BEAM POS. / FEEDBACK voltage ( [mV] , depending on
FEEDBACK OFF or ON ),
- the tidal gravity ( [mgal] ) and
- a weighting factor to be applied,
- preset to 1.0 and toggled by mouse clicks between
- 1. ( symbol = blue or magenta )
and
- 0. ( symbol = grey ).
- As a result of the digital simulation of the ( analog ) dial,
- in
general a nulling to exactly 0.0 [mV] is not possible.
- Counter and output voltage readings are therefore combined to
a raw reading ( [Counts] ),
- using the actual calibration factor ( the default value
DEF = 1.0, if not yet determined ).
- All meter readings ( except X LEV ADJUST )
presume a correct long level and cross level setting.
- For CALIBRATE OUTPUT / FEEDBACK
- the corresponding output voltages,
observed at constant counter setting, are the raw readings ( => no
calibration factor applied ).
- For X LEV ADJUST readings ( in
LABORATORY TEST ) observed at constant counter setting and
uniform setting of FBK ON / OFF,
- the result ( = position of the apex of a curve fitted to the
observations ) is almost independent of the calibration factor value.
- In order to process CALIBRATE COUNTER readings correctly,
- at least one of the voltage calibration factors is needed.
- For NETTLETON / MODEL PROFILE obsevations
- dial / counter and at least one voltage output have to be
calibrated.
- The obsevations can be displayed for different processing states :
- as raw readings ( RAW READ ... ) or
- after tidal and drift correction
( FIN READ ... ),
- with selectable abscissa :
- profile ( ... READ (PRO) ) or
- obsevation time ( ... READ (TIM) ).
- INSTR. DRIFT
- displays raw readings taken at the base station after tidal correction
- ( for X LEV ADJUST taken at
X LEV = 0 [seca], marked magenta ).
- A polynomial of max. second degree is fitted to the observed values and
- the resulting drift corrections is applied to all readings :
- => FIN =
RAW - tidal corr. - drift corr..
- RAW READ (TIM)
- displays readings after tidal correction, plotted versus obsrvation
time.
- Straight lines ( grey ) are drawn between multiple readings of
the same station
- to facilitate the "qality control".
- ! All lines should be parrallel to corresponding lines
between base station readings !
- FIN READ (TIM)
- displays readings after tidal and drift correction, plotted versus
time.
- Hoizontal straight lines ( grey ) are drawn through the mean
value of multiple readings of the same station,
- for base station readings the zero-order coefficient of the drift
polynomial is used.
- FIN READ (PRO)
- displays readings after tidal and drift correction , plotted
vesus station location.
- For X LEV ADJ ( LABORATORY TEST )
- a polynomal of second degree ( approximation of the cosine to be
expected )
- is fitted to the observed values, and
- the X LEV ZERO position
- is determined from the
apex of the curve.
- For CALIBRATE OUTPUT / FEEDBACK / COUNTER
- a straight line is fitted to the observations,
- and from its slope,
- the respective calibration factor is determined.
- For MODEL PROFILE
- a model curve is diplayed, calculated with the actual density value
- ( with DEF = 2.67 [g/cm**3], if not yet determined ).
- In the calibration procedures ( CALIBRATE ... )
- the respective calibration factors are calculated iteratively
( at least 2 iterations ),
- because their values are involved in the calculation of
tidal and drift corrections to be applied in advance.
- The parameters, determined in the various procedures,
- are displayed
( red ) at the upper right corner of the graphic display
- and
- are available for further calculations in the same program after clicking
USE
- ( holds as well for the user selected density values :
DEC / INC DENS in ... PROFILE ).
- Upper right corner blue :
- E = RMS error for curve fit to observations
- or
- COR TOP = correlation coefficient of topography
for BOUGUER ANOM. ( NETTLETON ).
- SAVE :
- ! Parameters have to be saved in order to be available to other
programs !
- CLEAR
- erases the readings,
- HIDE
- hides the window, to be restored ( without loss of data ) by
selecting the same procedure again.
! A program change erases all readings of the previous program !
Window READINGS
HowTo
Table of Content
Saving and Processing of Readings
- Meter readings are taken in RUN mode of the applet
( see Menue Panel ), and
- after adjustment of dial / counter to the gravity value of one
of the stations ( in general BAS ).
- Field Observations
- To take readings on profiles ( CALIB xxx, NETTLETON or
MODEL PROFILE ), the following loop has do be repeated for every
station :
- Lock meter ( "Black Lid" )
- Move to next station ( NEXT, PREV or BAS,
menue panel )
- Readjust L and X Level
( if not METER TILT : CONST.,
see window HELP )
- Unlock meter ( "Black Lid" )
- Adjust Dial / Counter
( "Black Lid", not for
CALIB. OUTPUT or FEEDBACK )
- Wait until the beam has reached its new equilibrum position
- ( use CHART RECORDER to observe beam velocity,
POS / FBK VEL )
- Save reading ( SAVE STN READING,
menue panel )
- X LEV ADJUST
- ( LABORATORY TEST, meter unlocked )
- NEXT, PREV and BAS ( menue panel ) are disabled and at fixed
setting of Dial / Counter, the above loop reduces to :
- Adjust X LEV to a new position preserving the
L LEV position ( "Black Lid" )
- Wait until the beam has reached its new equilibrum position
- ( use CHART RECORDER to observe beam velocity,
POS / FBK VEL )
- Save reading ( SAVE STN READING,
menue panel )
- If WARN ( window HELP ) is activated,
saving of obviously incorrect readings
( beam still moving, incorrect levelling ) is
inhibited.
- Processing and Interpretaion
- ( at least 3 readings saved : 2 x base station,
1 x profile, applet in RUN or HALT mode,
meter LOCKED or FREE )
- Saved readings are displayed as RAW READ (PRO) and
- before results of interpretation can be displayed
- in FIN READ (PRO) ( X LEV ADJ and
CALIBRATE xxx ) or
- in BOUGUER or FREE AIR ANOM.
( NETTELETON and MODEL PROFILE ),
- the display INSTR. DRIFT has to be called, because the resulting
drift correction
- has to be applied before calculating final results.
- In any display mode the weight of a reading
( def. = 1. ) can be toggeled between
1. / 0. using the left / right
mouse button :
- readings with weight 1. are marked +
( blue ),
- readings with weight 0. are marked x
( grey ),
- the display is switched back to RAW READ (PRO), and
- INSTR. DRIFT and interpretation displays have to be call again.
- Multiple readings of the same station are displayed separated
- in RAW / FIN READ (TIM) and
- thus are accessible individually.
- ( Enable INFO in window HELP, applet in HALT mode, to
identify the readings. )
- Final Results have to be saved
- ( SAVE in window READINGS => USER CALIB. and
ACTUAL VAL. in
window HELP )
- to be available in other procedures ( i.e. calibration
factors needed in field observations ).
Saving and Processing
HowTo
Table of Content
Window CHART RECORDER :
( 2-channel "analog" recorder, screenshot ) :
Example :
beam position and velocity after dial / counter change of -0.1 [Counts] =
approx. +100 [µgal]
Continuous display of output voltages, of their velocities,
of the tidal gravity and of the seismic noise.
Selectable in the menue panel of the recorder ( disabled
if the recorder is running ) :
- Signals to be displayed
-
- POS DIR and AVR
- beam postion unfiltered and average,
- FBK DIR and AVR
- feedback voltage unfiltered and average,
- ... VEL
- velocities of above signals,
- TIDAL G
- tidal gravity ( ampl 45 [µgal], start phase
random ),
- NOI VER and HOR
- vertical and horizontal component of seismic noise.
- Time scale of recording
- toggle switch FST ( fast ), SLW
( slow, 1 / 5 ).
- Sensitivity
- ±nn [V], [mV] or [V/sec], [mV/sec] for
full scale,
- ±50 [µgal] for TIDAL G,
- fixed values, depending on selected amplitude, for
NOI VER and HOR.
After a change of recording parameters the preceding signal is displayed
grey.
The output of the AVR low pass is displayed cyan during approx. 90 [sec]
after a manual or automatic reset ( to point to the settling time of the
low pass ).
Window CHART RECORDER
HowTo
Table of Content
Window CROSS COUPLING :
( screenshot )
Example :
mechan. sensitivity default ( L LEV 0 [seca] ),
FEEDBACK OFF
DIAL / COUNTER adjusted to BEAM POS 0 [mv]
harmonic acceleration, period 6.0 [sec],
amplitude vertical ±2 [mgal], horizontal
±30 [mgal], phase diff. 45 [Deg]
- Continuous display of
- horizontal and vertical acceleration
- ( X / Y display, individually scaled to resp. max. amp.,
- upper and right hand scale [mgal], red,
- act. val. = moving red point + red marks in resp. scale )
- and
- beam postion relative to mechanical stops
- ( left hand scale [seca], black,
- act. val. = moving red mark in scale ).
- Green :
- position of pivot, with horizontal straight line, indicating the
"True Horizontal" ( perpendicular to local
gravity )
- and
- the ideal position of the reading line
( green label xxx [seca] ) corresponding to the actual
L LEV setting.
- ( Here L LEV = 0 [seca] => pivot and green line at
0 [seca]. )
- Red arrows :
- components of the force acting on the center of gravity
- ( arrow legnth individually normalized to
2 ∗ mass radius ).
- Menue panel ( disabled if the acceleration signal is
switched on ) to adjust
- the amplitudes of vertical and horizotal acceleration,
- ( ver. : 0, 0.3 ... 10 [mgal],
hor. : 0, 0.3 ... 100 [mgal] )
- the phase difference between vertical and horizontal component
- ( DEC / INC PHS in selectable steps of
90, 15 and 1 [Deg] )
- and
- the period of the harmonic acceleration.
- ( 3, 6, 12, 24 and 48 [s] )
- O N starts / stops the harmonic acceleration signal
( COS - tapered at the beginning ).
Recorder display of Beam Postion Votage and corresponding
AVR low pass output
Note :
- A dc-offset of the beam postion voltage of approx.
+10 [mV],
- observed on the high gain display of the corresponding AVR low pass
output ( CHAN 2, ±20 [mV] F.S. ),
- would be interpreted as
- an aditiononal gravity change of approx. +10 [µgal].
- It corresponds to a beam position offset of approx. -1.5 [seca]
as a result of the horizontal force
- acting toward the pivot for mass positions below pivot
position ( "true horizontal" ),
- acting toward the scale for mass postions
above :
- negative ( downward ) excursion is increased,
- positive ( upward ) excursion is decreased
- => negative mean position => positive mean output
voltage.
- For given magnitudes of the vertical and horizotal component of
acceleration,
- the magnitude and the direction of the cross coupling
offset
- depends on the phase difference between beam excursion and
horizontal force,
- and thus on the phase difference between the two acceleration
components.
- For signal periods much longer than the the natural
period of the instrument,
- beam excursion and vertical acceleration are in phase,
- hence the cross coupling effect
- is greatest for a linear polarized accelerations
( 0 / 180 [Deg] ) and
- vanishes for circular polarisation
( +90 / -90 [Deg] ).
- In the example shown the signal period ( 6.0 [sec] )
is below the natural period of the meter
( > 12 [sec] ),
- resulting in a phase shift between vertical acceleraton
and beam excursion,
- and
- leading to a maximum cross coupling for approx. +45 and
-135 [Deg]
- ( almost no cross coupling for approx. 135 and -45 [Deg] ).
- Since the magnitude of the cross coupling effect strongly depends on the
magnitude of the beam excursion,
- cross coupling can be greatly reduced by activating the
feedback system and
- thus forcing the beam close to its zero position.
- A second consequence of feedback is the decrease of the
natural period ( here below 1 [s] ),
- leading to a almost vanishing cross coupling offset for a
driving signal
- of several seconds of period and
- a phase difference of approx. 90 [Deg] between
hor. and ver. component.
Example :
( FEEDBACK ON, harmonic acceleration, period 6.0 [sec],
amplitude vertical ±10 [mgal], horizontal
±30 [mgal], phase diff. 90 [Deg]
Recorder display of Feedback Votage and corresponding
AVR low pass output
The offset below 1 [mV] ( CHAN 2, FBK AVR )
corresponds to a cross coupling error below 1 [µgal].
- This is a demonstration of the cross coupling effect, physically
fairly correct, but with unrealistic driving signals :
- In shipborne gravity surveys "harmonic" vertical and horizontal
accelerations of up to 100 [mgal] at periods longer than
6.0 [sec] have to be handled.
- This affords
- mechanically less sensitive instruments and / or more powerfull feedback
systems
- ( and a different processing / filtering and recording of output
signals )
- to avoid mechanical clipping and to keep cross coupling errors within
tolerable limits.
Window CROSS COUPLING
HowTo
Table of Content
Examples of display, processing and interpretaion of readings
( Screenshots of Window "Gravity Meter READINGS" )
NETTLETON PROFILE
Raw readings along profile ( RAW READ (PRO) ) :
Raw readings versus observation time
( RAW READ (TIM) ) :
Instrument drift versus obsevation time
( INSTR. DRIFT ) :
After tidal and drift correction ( versus time,
FIN READ (TIM) ) :
After tidal and drift correction ( along profile,
FIN READ (PRO) ) :
Examples
HowTo
Table of Content
Bouguer anomaly ( BOUGUER ANOM. ) :
( calculated for density = DEF = 2.67 [g/cm**3] )
Free air anomaly ( FREE AIR ANOM. ) :
Free air anomaly / station height
( F. A. (STN HGT) ) :
( => density = 2.4000 [g/cm**3] )
Bouguer anomaly ( BOUGUER ANOM. ) :
( calculated for density = 2.4000 [g/cm**3] )
Bouguer anomaly ( BOUGUER ANOM. ) :
( calculated for density =
2.4000 + 0.0100 [g/cm**3] )
Bouguer anomaly ( BOUGUER ANOM. ) :
( calculated for density =
2.4000 - 0.0100 [g/cm**3] )
Examples
HowTo
Table of Content
ADJ X LEV ( LABORATORY TEST )
Interpretaion ( FIN READ (PRO) ) :
The position of the bubble, resulting in minimum sensitivity to
instrument tilt, corresponds to the apex of the red curve
( NEW ZERO MARK, red, upper right ).
By activating SAVE, the value can be saved as "USER CALIB."
( see Window HELP ).
CALIBRATE FEEDBACK
Interpretaion ( FIN READ (PRO) ) :
The calibration factor [mgal/V] for the output voltage FEEDBACK [mV]
corresponds to the reciprocal of the slope of the straigth line fitted to the readings
( NEW CAL FACT, red, upper right ).
By activating USE, the value can be transferred to text field
SCAL FACT : ... [mgal/V] for additional iterative
calculations of the coefficients.
By activating SAVE, the value in the text field can be saved as
"USR CALIB."
( see Window HELP ).
Examples
HowTo
Table of Content
Download
Class and html files for a local installation of the applet are available as
a
zip file and as
a
tar.gz file.
- These files do not include the PDF-file
( approx. 3 MB )
- L&R MODEL G AND D METER MANUAL.
- It is available for download at
L and R Meter Service
and at
Geophysics Dept.,
and
- it should be saved in the directory grav, where the
class- and html-files of the applet are stored after unpacking
the zip- / tar.gz-file.
More applets at :
the author's
Homepage
Rev. 15-feb-2007
Comments to
Fritz Keller
( ned gschempfd isch globd gnueg )
Top of Page
Table of Content
HowTo
Screenshots
Back to the
Applet Index ( Geophysics Dept.,
TU Clausthal )